This means calibrating shortly before the validation run (referred to as Pre Calibration) and then checking the accuracy afterwards to ensure that they are still within predefined tolerances (referred to as Post Calibration).Using the built-in calibration functionality, users can calibrate the Ellab sensors at regular intervals.Guidelines state that 11 temperature-measuring points are needed when performing a heat distribution study of an autoclave that is from 1.5 to 2.5 m3.This number is reached by placing one sensor in each of the eight corners in the autoclave, one in the center, one close to the drain and one next to the autoclave’s temperature sensor.Common sterilized products and containers are: An autoclave is a vessel that uses pressurized steam to heat the product to achieve sterilization.Autoclaves come in a wide variety of sizes, from small bench-top laboratory autoclaves to large production autoclaves.

In most cases it is necessary to monitor the temperature of the product and the pressure of the autoclave during sterilization.The latent heat released when steam condenses on the items to be sterilized makes this process very efficient.Sterilization temperatures range from 115 to 134°C.Commonly used standards and guidelines are ISO EN 285, ISO EN 17665, ISO EN 11138, HTM 2010 and PDA Technical Report No.1 Validation of Moist Heat Sterilization Processes.

steam and validating-67steam and validating-20steam and validating-85

This could be during batch production, cycle development and validation.